Objective. Students will interpret waveform capnography tracings and pulse oximetry readings to confirm correct airway placement, guide ventilation rate and volume, and recognize early signs of tube dislodgement or patient deterioration.
This lesson, its worked examples and its practice questions has not yet been verified by a licensed clinician in the field or by a certification or approval body. It is a study draft and may contain errors. Do not use it to make decisions about any patient or medication. Always follow your instructor, your supervising professional, and current official references and regulations.
VoTechWorks program approval and accreditation are still in progress. Credentials are issued only by the certifying bodies named in each program, after their own exams.
Once an advanced airway is placed, your job is not finished. Tube dislodgement, bronchospasm, and patient deterioration can occur at any moment during transport. Waveform capnography and pulse oximetry together give you a continuous, real-time picture of oxygenation and ventilation. Used correctly, they are among the most powerful tools in the paramedic's kit. Used without understanding their limitations, they can create false reassurance.
Capnography measures the concentration of carbon dioxide (CO₂) in exhaled air and displays it as a waveform over time. The numeric value displayed is the end-tidal CO₂ (EtCO₂), expressed in millimeters of mercury (mmHg). Normal EtCO₂ in a healthy adult at rest is roughly 35–45 mmHg, closely approximating arterial CO₂ levels.
The waveform itself has a recognizable rectangular shape with four phases:
A normal waveform has a flat alveolar plateau and returns cleanly to baseline with each breath.
Absent waveform (flat line): No CO₂ detected. In an intubated patient this strongly suggests esophageal intubation or complete tube obstruction. Treat as misplacement until proven otherwise — immediately confirm tube position clinically and with direct laryngoscopy.
Hyperventilation: Normal waveform shape but EtCO₂ below 35 mmHg. Ventilation rate or tidal volume is too high. Slow the rate and check tidal volume.
Hypoventilation: Normal waveform shape but EtCO₂ above 45 mmHg. Ventilation is inadequate. Increase rate or volume as guided by protocol.
Shark-fin pattern: The alveolar plateau slopes upward rather than remaining flat, creating a pointed peak. This indicates obstructive airflow, most commonly bronchospasm or reactive airway disease. Administer bronchodilator therapy per protocol.
Cardiac arrest waveform: Very low EtCO₂ (often below 10 mmHg) with small, irregular waveforms. EtCO₂ reflects pulmonary perfusion; low values in arrest reflect poor cardiac output. A sudden sustained rise in EtCO₂ during CPR may signal return of spontaneous circulation (ROSC).
Gradually declining EtCO₂: Consider tube displacement (partial), decreasing perfusion, or a leak in the ventilation circuit. Reassess immediately.
Continuous waveform capnography is the gold standard for confirming and monitoring endotracheal tube placement in the prehospital environment. A consistent, rectangular waveform with appropriate EtCO₂ values across multiple breaths confirms the tube is in the trachea. Colorimetric CO₂ detectors change color but do not provide the ongoing monitoring that a waveform device does. Any time the waveform disappears or becomes erratic, treat it as a tube problem until you rule one out.
Pulse oximetry (SpO₂) measures the percentage of hemoglobin saturated with oxygen using light absorption. A reading of 95% or higher is generally considered acceptable in most patients, though clinical context always applies.
SpO₂ is fast and non-invasive, but it has important limitations every paramedic must know:
Because of these limitations, SpO₂ must always be interpreted alongside capnography, clinical assessment, and patient history. Never withhold oxygen from a symptomatic patient because SpO₂ appears normal.
Think of capnography and pulse oximetry as two different windows into the same patient. Capnography tells you about ventilation (CO₂ elimination) and tube position in near-real time — changes appear within a breath or two. Pulse oximetry tells you about oxygenation, but it lags behind ventilation problems by seconds to minutes because of oxygen reserves already in the blood. A patient who loses their airway will show a capnography change before the SpO₂ drops. This lag makes capnography the earlier warning system.
Document EtCO₂ and SpO₂ values at regular intervals on your patient care report. Trending matters: a gradual rise in EtCO₂ or a slow fall in SpO₂ may be more significant than a single abnormal reading.
Answer each question, then check it. Nothing is recorded; this is your own practice.
A free tech program — complimentary, as our commitment to get America working. Help us build VoTechWorks. Tell us what works, what doesn’t, and what you want to learn.
Does the school offer financial aid?
No. We don’t offer financial aid because we’re free. No, really: completely free.
Do you offer industry certification?
No. We are not certified yet, so we can’t offer that to students. What we do offer is knowledge and understanding of a trade.
Enroll first in the trade you’re interested in. You get your key, your program opens, and you begin studying.
Enroll free →